en
Global - english
Sun Sep 24 23:16:03 2017
Sun Sep 24 23:19:38 2017
ISOMETER® 107TD47

ISOMETER® 107TD47

The ISOMETERs® of the 107TD47 series monitor the insulation resistance of single and three-phase AC IT systems in medical locations. In addition, the IT system transformer is monitored for overload and overtemperature. In combination with the alarm indicator and test combination MK2430 they particularly comply with the requirements of DIN VDE 0100-710 (VDE 0100 Teil 710) and IEC 60364-7-710.

Do you already know our new product generation:

  • Features
  • Applications
  • Accessories
  • Ordering information
  • Downloads
  • Monitoring device for medical IT systems AC, 3(N)AC
  • Adjustable response value 50…500 kΩ
  • Load and temperature monitoring
  • Alarm LED
  • Monitoring of the connection to the system, earth, measuring current transformer, temperature sensors
  • Test button
  • External test button can be connected
  • BMS-bus interface
  • Common alarm relay with one potentialfree changeover contact
  • Backlit LC display
  • IT systems for the power supply in medical locations, hospitals, medical practices and outpatient surgical centres
  • IT systems with restricted power source with additional load current monitoring

 

Optional accessories for ISOMETER® 107TD47


TypeSupply voltage US (V)Art. No.
STW2 (Measuring current transformer)B942709
AN450 (Power supply unit)AC 230 VB924201
AN450-133 (Power supply unit)AC 127 VB924203
CMS460-D4-2 (Three phase loads)AC 70…276 V, 42…460 HzB94053030

Variants of ISOMETER® 107TD47


TypeNominal system voltage Un (V)Supply voltage US (V)Art. No.
107TD47AC 230 V, 50…60 HzAC 230 V, 50…60 HzB92016003
107TD47-133AC 127 V, 50…60 HzAC 127 V, 50…60 HzB92016004

Downloads for ISOMETER® 107TD47

Category

Language

Category Product Language Date Size Download